Download PAS (Link to Canon medical global website) Download the manual for applying PAS
Axial DWI with Exsper
| Sequence | SEEPI2D |
| Parallel Acquisition | Exsper 2x1 |
| Scan time | 0:38 |
| Resolution | 1.6x1.6 mm |
| FOV | 23 cm |
| Matrix size | 144x144 |
| Slice thickness | 5 mm |
| Slice gap | 1 mm |
| Number of slices | 25 |
| TR | 4130 ms |
| TE | 77 ms |
| NAQ | 1 |
| b-value | 1000 |
Axial FLAIR with Exsper
| Sequence | FSE2D |
| Parallel Acquisition | Exsper 2x1 |
| Scan time | 1:40 |
| Resolution | 1.0x1.0 mm |
| FOV | 23 cm |
| Matrix size | 224x224 |
| Slice thickness | 5 mm |
| Slice gap | 1 mm |
| Number of slices | 25 |
| TR | 10000 ms |
| TE | 104 ms |
| NAQ | 1 |
Axial TOF MRA with SPEEDER, Fast 3D mode 50%
| Sequence | MRA3D |
| Parallel Acquisition | SPEEDER 2x1 |
| Scan time | 1:15 |
| Resolution | 1.3x0.8 mm |
| FOV | 20 cm |
| Matrix size | 160x256 |
| Slice thickness | 1.3 mm |
| Number of slabs | 3 |
| Number of slices | 116 |
| TR | 14 ms |
| TE | 6.5 ms |
| NAQ | 1 |
Axial T2*w with Exsper
| Sequence | FE2D |
| Parallel Acquisition | Exsper 2x1 |
| Scan time | 0:53 |
| Resolution | 1.0x1.0 mm |
| FOV | 23 cm |
| Matrix size | 240x240 |
| Slice thickness | 5 mm |
| Slice gap | 1 mm |
| Number of slices | 25 |
| TR | 380 ms |
| TE | 9 ms |
| NAQ | 1 |
Axial T2w with Exsper
| Sequence | FSE2D |
| Parallel Acquisition | Exsper 2x1 |
| Scan time | 0:40 |
| Resolution | 0.9x0.9 mm |
| FOV | 23 cm |
| Matrix size | 256x256 |
| Slice thickness | 5 mm |
| Slice gap | 1 mm |
| Number of slices | 25 |
| TR | 4400 ms |
| TE | 90 ms |
| NAQ | 1 |
Axial T1w with Exsper
| Sequence | SE2D |
| Parallel Acquisition | Exsper 2x1 |
| Scan time | 0:56 |
| Resolution | 1.0x1.0 mm |
| FOV | 23 cm |
| Matrix size | 224x224 |
| Slice thickness | 5 mm |
| Slice gap | 1 mm |
| Number of slices | 25 |
| TR | 428 ms |
| TE | 8 ms |
| NAQ | 1 |
Coil
Octave SPEEDER Head 1.5T
Download PAS (Link to Canon medical global website) Download the manual for applying PAS
| Sequence | FASE3D |
| Parallel Acquisition | SPEEDER 2x1 |
| Scan time | 5:51 |
| Resolution | 1.3x1.3 mm |
| FOV | 43x34 cm |
| Matrix size | 336x256 |
| Slice thickness | 2 mm |
| Number of slices | 50 |
| TR | 3250 |
| TE | 369 |
| NAQ | 2 |
Coil
Octave SPEEDER Head 1.5T
Atlas SPEEDER Body 1.5T
Octave SPEEDER Spine 1.5T
Download PAS (Link to Canon medical global website) Download the manual for applying PAS
| Sequence | FASE2D |
| Parallel Acquisition | Exsper 2x1 |
| Scan time | 0:03 |
| Resolution | 1.5x1.1 mm |
| FOV | 38x28 cm |
| Matrix size | 256x256 |
| Slice thickness | 60 mm |
| Number of slices | 1 |
| TR | 3000 |
| TE | 1000 |
| NAQ | 1 |
| Sequence | FASE3D |
| Parallel Acquisition | SPEEDER 2x1 |
| Scan time | 0:20 |
| Resolution | 1.0x1.0 mm |
| FOV | 32x25.6 cm |
| Matrix size | 320x256 |
| Slice thickness | 3.5 mm |
| Number of slices | 20 |
| TR | 2850 |
| TE | 800 |
| NAQ | 1 |
Coil
Atlas SPEEDER Body 1.5T
Octave SPEEDER Spine 1.5T
© Canon Medikal Sistemler Turkey A.Ş.
© Canon Medikal Sistemler Turkey A.Ş.